iMETOS® 3.3 MANUAL: CONTENT
Welcome among iMETOS 3.3 users!
Thank you for choosing an iMETOS for monitoring soil moisture or environmental data. The iMETOS 3.3 has been designed to monitor data with wide variety sensor sets. Like all products of the iMETOS family it measures, logs and sends the data to the Internet.
iMETOS 3.3 systems will be mainly used for:
- Microclimatic data monitoring for Farming decision.
- Soil moisture monitoring with volumetric or tensiometric sensors.
- Irrigation monitoring with flowmeter, pressure switch, tube pressure, pH and E.C.
- Frost alarms via SMS with 3 inputs for high precision PT1000 temperature sensors: Air, Dry bulb and Wet bulb.
- Rain, flood & snow monitoring with rain gauge, pressure transducers and ultrasonic sensors.
- Disease models forecasting with basic sensor set such as leaf wetness, temperature and relative humidity sensors, rain gauge.
- Hyper localized weather forecast corrected with local measurements.
Among all the many new features you will find:
- Firmware update “over the air”
- Real Time Operating System (RTOS)
- Extended data and program memory
- On-board modem for GSM, CMA, UTMS
- Self location with on-board GPS
- Native USB port
- Access point for remote wireless sensor node
FAQ
1. YOUR iMETOS® 3.3
1.1 iMETOS® 3.3 BOARD DESCRIPTION
1.2 PROCESSOR
1.3 MODEM
1.3. 1 LTE (4G) VERSION
1.3. 2 UMTS (3G) VERSION
1.4 EXTERNAL FLASH MEMORY
1.5 POWER SUPPLY
1.6 SIM CARD HOLDER
1.6. 1 HOW TO DEACTIVATE PIN REQUEST
1.6. 2 USING UNUSUAL SIM CARDS
1.7 USB PORT
1.8 JUMPERS
1.9 LEDS AND BLINKING CODE
1.10 TEST POINTS
1.11 SENSOR INPUTS
1.11. 1 SENSOR IDENTIFICATION
2. START UP YOUR iMETOS®
2.1 USE YOUR iMETOS®
2.2 REGISTER YOURSELF AS A USER ON FieldClimate.com
2.3 ADD THE iMETOS® TO YOUR ACCOUNT
3. INSTALLING THE iMETOS®
3.1 INSTALLING CLIMATE SENSORS
3.2 INSTALLING SOIL MOISTURE SENSORS
3.3 INSTALLING TENSIOMETERS
3.3. 1 FILLING AND RE-FILLING THE TENSIOMETER
3.4 INSTALLING WATERMARK SENSORS
3.5 INSTALLING DRILL AND DROP PROBES
3.6 WIRING A JOHN DEERE PROBE
3.7 WIRING AN AQUACHECK PROBE
3.8 iMETOS® 3.3 RADIO ACCESS POINT
3.9 SETUP YOUR iMETOS® WiFi NETWORK
4. INSTALLING NEW SENSORS ON THE iMETOS®
4.1 INSTALLING AIR, SOIL, WET OR DRY BULB TEMPERATURE
4.2 INSTALLING A GLOBAL RADIATION SENSOR
4.3 INSTALLING A WIND SENSOR
4.3. 1 PI ULTRASONIC WIND MONITOR
4.3. 2 WIND DIRECTION SENSOR
4.4 INSTALLING BAROMETRIC PRESSURE SENSOR
5. MAINTAINING THE iMETOS®
5.1 MAINTAINING THE RAIN GAUGE
5.2 MAINTAINING THE LEAF WETNESS SENSOR
5.3 MAINTAINING THE TEMPERATURE AND RELATIVE HUMIDITY SENSOR
5.4 MAINTAINING THE SOIL TEMPERATURE SENSOR
5.5 MAINTAINING THE GLOBAL RADIATION SENSOR
5.6 MAINTAINING THE WIND SPEED SENSOR
6. FIRMWARE
6.1 UPDATE OVER THE AIR
6.2 STATION WORKING MODES
6.3 EVENTS
7. WEB SERVER COMMUNICATION
7.1 STATION TO WEB-SERVER COMUNICATION
7.2 WEB-SERVER RESPONSE
7.3 USER DEFINED SETTINGS
7.3. 1 RAIN MONITORING
7.3. 2 WATER LEVEL MONITORING
7.3. 3 MEASUREMENT AND LOGGING INTERVAL
7.3. 4 FIXED TRANSFER INTERVAL AND TRANSFER SCHEDULE
7.3. 5 EMERGENCY SMS
7.3. 6 SMS WARNING SYSTEM
8. SETTING COMMUNICATION PARAMETERS AND RESETTING iMETOS® VIA SMS
9. USB COMMUNICATION (PI FIRMWARE UPLOADER)
9.1 USB COMMUNICATIONS MODES
9.2 USING THE TERMINAL MODE
9.3 TESTING SENSORS
9.4 UPLOADING FIRMWARE
9.5 CHANGING THE SERIAL NUMBER
10. APENDIX I.
11. APENDIX II.
FAQ
There aren’t any so far. If you encounter any problems with your iMETOS® 3.3, contact our support team at: support@metos.at.
1. YOUR iMETOS® 3.3
1.1 iMETOS® 3.3 BOARD DESCRIPTION 1.2 PROCESSOR 1.3 MODEM 1.4 EXTERNAL FLASH MEMORY 1.5 POWER SUPPLY 1.6 SIM CARD HOLDER 1.7 USB PORT 1.8 JUMPERS 1.9 LEDS AND BLINKING POINT 1.10 TEST POINTS 1.11 SENSOR INPUTS
The basic iMETOS consists of one stainless steel holder with the box containing the electronics, the battery, the solar panel and the dual antenna. In the package with the iMETOS you will also find two clamps to mount it on the pole. If you ordered a wind speed sensor you will find the three-cup wheel and the allen key to fix it to the axis of the sensor; if you also ordered the wind direction sensor, you will find a ”T” shaped holder with two clamps. Taped to the solar panel you will find a label with the unique serial number and the keys that enable the access to the FieldClimate platform.
You will find the battery mounted inside the station. The rain gauge is mounted on the right side of the station. The solar panel, the radiation shield with temperature and relative humidity sensor and the global radiation sensor will be connected with the main electronic box and have to be mounted on the holder.
Other sensors, like wind speed, wind direction, leaf wetness, soil temperature, wet and dry bulb temperature you will find connected to the main electronic box with 3m or 5m long cables.

1. Temperature and relative humidity sensor with radiation shield; 2. Global radiation sensor; 3. Dual antenna (GPS/communication); 4. Rain gauge; 5. Logger and modem; 6. Power supply (solar panel and battery); 7. Wind speed sensor.
The selection of sensors, connected to your iMETOS, depends on the model you have ordered as below.

1.1 iMETOS® 3.3 BOARD DESCRIPTION
Interface:
A) Extension board connector iMETOS® 3 1ozn
B) LED indicators
C) Connect button
D) Battery connector
E) Solar panel connector
F) Jumper 1
G) Jumper 2
H) USB connector
I) Boot jumper
J) Reset button
K) Antenna connecto
Picture 2: iMETOS 3.3 board – interface
Sensor inputs:
1) WIND iMETOS® 3 2ozn
2) SWITCH
3) LEAF
4) INPUT 2
5) INPUT 3
6) INPUT 4
7) REED
8) INPUT 1
9) RAIN
10) HC2 B
11) HC2 A
12) Direct PIbus (chain) input
Picture 3: iMETOS 3.3 board – sensor inputs.
1.2 PROCESSOR
• 32bit ARM cortex M3 processor (Energy Micro EFM32)
• Real time operating system (RTOS)
1.3 MODEM
1.3. 1 LTE (4G) VERSION 1.3. 2 UMTS (3G) VERSION
1.3. 1 LTE (4G) VERSION
Sierra Wireless® AirPrime® HL embedded modules offer unprecedented scalability between 2G/3G/4G, low-power consumption with up to 10Mbps download data speed.
One of the following modem is mounted on the socket of the board, as each region has different network parameters:
- HL7692: For EMEA region (Europe). It is a 4G-LTE Cat-1 module with fallback to 2G-GSM/GPRS/EDGE. Supported frequency bands are 4G-LTE B3, B8, B20 and 2G-900/1800
- HL7618: For North America region (USA – Verizon carrier). It supports 4G LTE frequency bands B4 and B13.
- HL7688: For North America Region (Canada – AT&T carrier) with fallback to 3G. Supported frequency bands are: 4G-LTE B2, B4, B5, B17 and 3G-B2/B5.

1.3. 2 UMTS (3G) VERSION
- For GSM/GPRS/EDGE/3G, i.e G2 networks:
Sierra Wireless AirPrime SL808x Series a self-contained E-GSM/DCS/GSM850/PCS-GPRS/EGPRS 900/1800/850/1900/ WCDMA 800/860/900/1900/2100 quad-band module. SIM cardholder. - For CDMA/EVDO networks (U.S. of America): Cinterion PCS3. Not SIM cardholder.
- GPS included
1.4 EXTERNAL FLASH MEMORY
The external flash memory is an Adesto AT45DB641E (8MB).
The memory is split into several blocks:
• 6MB reserved for weather data in Base64 format (i.e. the measured data), implemented as circulating memory (i.e. when memory is full, the oldest data is overwritten with newer data)
• 1MB reserved space for firmware update over the air
• 1MB reserved for configuration, performance and behaviour, including:
• Station settings
• APN tables (Acess Point Names with MCC, MNC, username and password)
• Backup (serial number and similar important settings that normally are not changed)
• Sensor image stores the configuration of the set of sensors connected to the iMETOS® 3.3. This info will be re-checked automatically every day at midnight so newly connected sensors will be detected and included in the logs. Manual update of this part of the memory can be done by pressing the reset button (J in picture 1)
• Event history (information about communication network and internal parameters, detailed description is given later on in this manual)
1.5 POWER SUPPLY
The 6V-4Ah battery is connected to the iMETOS® 3.3 board in the specific connector (D in the picture 1), the solar panel is connected to the solar panel connector (E in the picture 1).
Both power inputs are monitored by the iMETOS® 3.3 in the same way as the rest of the sensors.
Note: jumper on pins between D and E connector is not needed in last board revisions. The iMETOS® 3.3 will automatically change operational mode according with the charge level of the battery following these rules:
a) If battery drops to 6V, data logging continues, but data transmission quits until battery recovers
adequate charge level supplied from the solar panel.
b) If battery recovers the voltage value of 6,1V (before reaching the threshold of 5,3V), iMETOS® 3.3
restarts data transmission without manual intervention.
c) If battery reaches the limit of 5.3V iMETOS® 3.3 enters sleep mode and also stops data logging. To escape from sleep mode manual reset will be needed after the battery recovers the proper charge level.

1.6 SIM CARD HOLDER
1.6 1 UPDATE OVER THE AIR 1.6. 2 STATION WORKING MODES
In GPRS, UTMS, LTE (i.e. GSM G2, G3, G4) networks a SIM card from a provider is needed. To insert or remove a SIM card:
1. Slide the metal part towards the left to unlock the holder (see picture below)
2. Open the holder from the right side and insert or remove the SIM card
3. Close the holder
4. Lock the holder by sliding the metal towards the right.

1.6. 1 HOW TO DEACTIVATE PIN REQUEST
This can be done using:
• a GSM handset device or
• directly with the USB<=>PC connection (From Firmware version 4.0 and later, it is possible to do it via terminal menu. Find a detailed description in the section USB communication)
1.6. 2 USING UNUSUAL SIM CARDS
We have prepared an extanded table of Internet connection settings for various cellular service providers worldwide, which is present in the device memory (we are adding new providers all the time). You can check if your provider is in the table with sending email to: support@metos.at
It might happen that your iMETOS® is one of the first devices set up with a specific provider. In this case, you will need to set these parameters with your PC via USB port or sending SMS with correct settings to the station:
To set new APN settings send a special SMS to the station, with APN settings, its username and its password. Before sending the SMS insert SIM card into your device. The SMS with APN settings has the following form:

1.7 USB PORT
Direct communication PC <=> iMETOS® 3.3 is possible with a standard USB/mini-USB cable, using any terminal program. On below picture you can see a mini-USB connector on iMETOS® 3.3 board.

Note:
For non-Windows users:
• Drivers: www.silabs.com
For Windows users:
• Drivers and Terminal at: www.metos.at
For iMETOS® 3.3 Pessl Instruments has develop an specific terminal program named “PI Firmware uploader” witch combines on site firmware updating capabilities with serial terminal connection to the iME®TOS 3.3, you can find it at: www.metos.at
Via USB connection user can perform the following tasks on-site
1. Fully update the iMETOS® firmware
2. Check all sensors
3. Get report and events from the unit
4. Follow detailed communication process
5. Retrieve stored data in binary form
6. From Firmware version 4.0 and later check network signal strength and unlock SIM card
A detailed description of USB communication is given in the section USB communication with PC.
1.8 JUMPERS
There are 4 sets of open pair of pin in the iMETOS® 3.3 mother board (see below picture).

Jumper 1: on the board it is labeled as “J1”. When you put jumper on this pins, you will be able to access terminal menu of the board with your PC.
Jumper 2: on the board it is labeled as “J2”. When you put jumper on this pins, you will be able to access the modem through terminal directly. This is intended and used only by iMETOS® developers.
Boot jumper: on the board it is labeled as “BOOT”. Jumper on this pins is needed when you want to manually upload new firmware to the board with PI Firmware Up-Loader application. More on this is available in the Uploading firmware page.
Warning!!
Jumper 1 and jumper 2 function is conditioned to the connection of the USB cable. Without cable they are ignored. This is not the case of the BOOT jumper if you forget to remove this jumper the station will not go into normal mode of operation and you will have to return to the station and fix this inconvenience manually.
1.9 LEDS AND BLINKING CODE
iMETOS® 3.3 has a set of 3 LEDs on top left on the motherboard and an additional LED under the SIM card holder. The top left row gives information about the running processes, and the stand-alone LED indicates the operation of a modem (see picture below).

Important note:
Complete monitoring of the communication process should be done after every installation by connecting the PC to the iMETOS® 3.3. How to do this is described in detail on page USB communication with PC.
Modem (stand alone LED just under SIM card holder):
• Green LED: on, during the connection process to the GSM network, blinking slowly (short on and long off) when it is connected to the GSM network.
Processor (from left to right, top left row of LEDs):
• Green LED (left): USB cable is plugged (except when in boot mode – i.e. the boot jumper I is plugged on BOOT pins)
• Yellow LED (center): indicates one of the following:
1. In boot mode (plugged jumper): Pessl bootloader is installed 2. in terminal mode (plugged jumper 1) the meaning of the LED is given on the screen of the terminal program in the PC3. 3. in normal/scheduling mode (without any jumpers):
a) if the modem is on: modem blink code:i. short on – long off: waiting for modem, network found, SIM card is active
on――――on――――on――――on――――on――――on――――on――――on――――on――――on
ii. short on – short on – long off: connected to GSM
on-on――――on-on――――on-on――――on-on――――on-on――――on-on――――on-on
iii. short on – short on – short on – long off: connected to FieldClimate
on-on-on――――on-on-on――――on-on-on――――on-on-on――――on-on-on――――on-on-on
iv. short on – short on – short on – short on – long off: data has been sent and commands from the server have been received.
on-on-on-on――――on-on-on-on――――on-on-on-on――――on-on-on-on――――on-on-on-on
b) If the modem is off: the measurement process is in progress, it will light a few seconds every 5 minutes.
Tip:
Put all the attention in the center LED of the second row if you see the sequence of four blinks then you will know that the upload has been successful, if the sequence is not completed the red LED aside will light the take note of the last blinking mode before to identify the error.
Red LED (right):
1. communication error (if modem is on)
2. serious system error (if modem is off)
In both cases you will need to connect the iMETOS® 3.3 to the PC and use PI Firmware uploader to make the diagnostic and extract useful information to send to support@metos.at
1.10 TEST POINTS
In the iMETOS® 3.3 board there are several test points for convenience during service, they let the user access different voltages with no need to unplug or removing anything.

BAT: battery voltage
SOL: solar panel voltage
VSUP: internal voltage supply 6V
V31: internal voltage supply 3.1V (processor, hygroclip sensors…)
V54: internal voltage supply 5.4V (sensor inputs…)
Example: if you wish to measure battery voltage, you can measure it with Voltmeter on GND and BAT test points.
1.11 SENSOR INPUTS
There are 12 direct sensor inputs on the iMETOS® 3.3 board, supporting up to 600 sensors units.
Main sensors:
Some sensors have to be connected to specific inputs dedicated to them:
• WIND: wind speed sensor
• LEAF: leaf wetness sensor
• RAIN: rain gauge
• HC2 (two inputs): hygroclip sensors (air temperature and relative humidity)
• SWITCH input: any switch-type sensor can be connected (e.g. another leaf wetness sensor, tilt sensor, a presostat…).
• REED input: any reed-type sensor can be connected (e.g. another rain gauge, flow counter, tilt sensor …)
Sensors connected to these inputs as well as the battery and solar panel voltage measurement are referred to as “main” sensors.
INPUT 1-4:
INPUT 1-4 can be used for connecting either
1. Individual sensors (any of the complete range of PI_duty_sensors: Global radiation, temperature, wind direction, barometric pressure, water level, weighting cells….)
2. PI_bus chains of modules with sensors.
On these inputs, the type of a connected sensor (or whether a chain is connected) is automatically recognized based on Pessl Instruments proprietary chain protocol.
Different temperature sensors are based on the same IC and consequently the iMETOS® 3.3 does not distinguish among them. Therefore default inputs have been reserved for each of this temperature sensors as follows:
• INPUT 1: air temperature
• INPUT 2: dry bulb temperature
• INPUT 3: wet bulb temperature
• INPUT 4: soil temperature
Example:
If a dry bulb temperature happened to be connected to input 4 it will receive the name of “soil temperature” but this will be only the default denomination and the user can change it later without further consequences on FieldClimate portal (This is then at your Account in FieldClimate.com / iMetos settings / names).
Sensors connected directly to the INPUT 1-4 are referred to as “duty-cycle” sensors because data is transmitted by means of the pulse width. Examples are temperature, global radiation, wind direction, and barometric pressure (a complete list is on appendix I).
A PI_bus chain is recognized by its characteristic frequency of 750 Hz. Sensors of the chain are recognized by identifiers (see next page Sensor Identification).In addition to the possible chain inputs on INPUT 1-4, the iMETOS® 3.3 has one dedicated input for chain.
Warning!!
To connect a chain to INPUT 1-4 a 485-to-duty-cycle converter is needed in the connector of the bus (as in previous iMETOS® boards), the bus cable can’t be connected directly.
Direct chain input: In addition to the possible chain inputs on INPUT 1-4, the iMETOS® 3.3 has one dedicated input for every chain.
Note: The PI_bus chain input has no need of any 485 to duty converter, the bus cable is connected directly.
Extension board connector:
To the extension connector one of the following extensions can be connected:
- Aquacheck for soil moisture probes (not John Deere probes) and with the necessary changes for ultrasonic wind speed sensors (only channel A)
- SDI12 – 1 input for 1 soil moisture probe
- SDI12 – 2 inputs for soil moisture probes and ultrasonic wind speed sensor (only channel A)
- Modbus: for ultrasonic snow depth sensors
- RS485: two more PI-Bus channels
- Radio nodes AP Board for interfacing the iMetos33 station with radio nodes stations

1.11. 1 SENSOR IDENTIFICATION
MAC-Address:
Every sensor has to have a different proper and unique identification (unique ID), this is the MAC address or chain channel number (assigned by iMETOS®).
Chain_channel number:
The chain_channel number identifies the position of a sensor on the iMETOS® 3.3. It is made up of the chain number and the channel number of the sensor.
The chain number is assigned according to the input connector:
Chain 0 = Main sensors + HC2 + Duty sensors
Chain 1 = Chain at input 1
Chain 2 = Chain at input 2
Chain 3 = Chain at input 3
Chain 4 = Chain at input 4
Chain 5 = Chain at dedicated input (RS485)
Chain 6 = Extension board input A / Radio node sensors
Chain 7 = Extension board input B
Chain 8 = Radio node sensors
The channel number of any sensor connected to the PIbus is assigned according to the position of the sensor on the chain node and the proximity of the node to the iMETOS®. For RadioNode is assigned in chronological order of peering with the iMETOS®. The channel numbers of Chain 0 are given in the following table:
The chain_channel number is then calculated as Chain number * 2000 + Channel number
Examples:
a) Direct connected sensors
Wind speed sensor connected to WIND input: Chain = 0, Channel = 6 → Chain_channel = 6 (as result of: 0·2000+6=6)
Global radiaton sensor connected in input sensor connected to input_1: Chain = 0, Channel = 1 → Chain_channel = 1 (as result of: 0·2000+1=1)
b) Pibus conneted sensors
Last (*) sensor in a PIbus chain with 4 sensors connected to input 3:
• Chain = 3, Channel = 4 → Chain_channel = 6004 (as result of: 3·2000+4=6004)
Third (*) sensor in a PIbus chain with 18 sensors connected to dedicated input:
• Chain = 5, Channel = 3 → Chain_channel = 10003 (as result of: 5·2000+3=10003)
Sensor code:
The sensor code specifies the sensor type, in Appendix 1 one you will find the complete list at the time of this edition. An example of codes are:

2. START UP YOUR iMETOS®
2.1 USE YOUR iMETOS® 2.2 REGISTER YOURSELF AS A USER ON FieldClimate STATION WORKING MODES 2.3 ADD THE iMETOS® TO YOUR ACCOUNT
2.1 USE YOUR iMETOS®
FieldClimate is the web service you are intended to use your iMETOS® with. It allows you to see the data in graphs or tables. It provides interfaces for automised downloads and it provides a powerful decision support system for plant protection and irrigation.
2.2 REGISTER YOURSELF AS A USER ON FieldClimate.com
To use the services on FieldClimate it is necessary to register as a user first. Click the plus (+) button to add a new user.
The registration screen which comes up, asks you for a username and a password and it needs your email address as well as your postal address and some information about the company. Please note you will have to enter the real email address. An acknowledge email is send to your inbox and its containing link that has to be used to activate the newly created user account on FieldClimate.

2.3 ADD THE iMETOS® TO YOUR ACCOUNT
If you have got the activation email and you activated the account you will be able to enter to FieldClimate. To add your device, open user menu and then press the “Add/Remove station” button.


3. INSTALLING THE iMETOS®
3.1 INSTALLING CLIMATE SENSORS 3.2 INSTALLING SOIL MOISTURE SENSORS 3.3 INSTALLING TENSIOMETERS3.4 INSTALLING WATERMARK SENSORS 3.5 INSTALLING DRILL AND DROP PROBES 3.6 WIRING A JOHN DEERE PROBE 3.7 WIRING AN AQUACHECK PROBE 3.8 iMETOS® 3.3 RADIO ACCESS POINT 3.9 SETUP YOUR iMETOS® WiFi NETWORK
3.1 INSTALLING CLIMATE SENSORS
Each iMETOS® can be expanded to the maximum possible number of sensors. It is equipped with specific inputs for rain, wind speed, leaf wetness, temperature, and relative humidity and for soil moisture probes. Beside this it has 4 digital inputs more which can be used for a range of different sensors and sensor chains.
iMETOS® is able to distinguish between the different sensors added to the 4 digital inputs. If we connect global radiation, a photosynthetic active radiant sensor, a wind direction, a barometric pressure sensor or a temperature sensor – they can be detected by their frequency.
• Before connecting a NEW SENSOR to the motherboard, unplug battery and solar panel. Once connected, plug them in again.
• It is recommended to do always a SENSOR TEST before leaving the site to ensure the station reads properly the new sensor.
Solar panel, battery, temperature and relative humidity, global radiation, rain gauge and barometric pressure are part of the main system. There are usually no long cables for these sensors. Wind speed and Wind direction sensors will need a cable because they have to be mounted on the end of the pole. They normally come with 3 m cable. The soil temperature sensor comes with 5 m cable. It has to go to the ground at the desired depth. Leaf wetness comes with 5 m cable. This sensor is mounted in the canopies of trees and vines.
Convection cup and global radiation sensor are mounted on top of the holder. They are dismounted during transportation and have to be mounted after unpacking. Just push them under the 3 screws which are prepared in the housing for each of them and fix the screws with a small screwdriver.
The solar panel might be clapped down for transportation. Screw the 3 screws in and it will be clapped out. On some systems for cooler climate the solar panel can come on a separate holder. In this case please fix this holder on the mounting pole by using the clamps which have been added to the package for this purpose.
The temperature and relative humidity sensor should be mounted between 1.25 m and 2 m above the ground depending on the crop. This is the standard measurement for these climate factors in agricultural meteorology.
The iMETOS® is mounted on the post with two clamps. Make sure that the pole is buried as straight as possible. How horizontal the climate station is mounted can be checked by the little water lens at the rain gauge.
On northern hemisphere the solar panel should be exposed to south and the same the global radiation sensor. The wind direction sensor should point towards north and the wind speed sensor should point towards south again.
The leaf wetness sensor comes with a cable tie to be mounted on a branch of the tree or vine or on the main holder (next to the rain gauge) slightly inclinated. Please mount this sensor in a position with allows the sensor to pick up the early rain and to stay wet in shadow.
The soil temperature sensor should be buried in the ground. How deep it should be buried is depending from the goal which should be reached with the sensor. To have an impression of root growth and nitrogen mineralization in vines or apples in early spring it is best be mounted at 10 to 20 cm. It you want to assess the emergence possibilities of seeds it is depending on the sowing depth of the seed. Using it in corn (maize) would mean to mount it at 5 cm and using it for sugar beets would mean to mount it at 2 or 3 cm only.

Water level sensors are equipped with 15 m cable to give the sensor into the water. It is combined with a barometric pressure sensor.

Installing PI-bus sensors
PI-bus sensors (Tensiometer, Water level, etc.) can be directly connected to RS485 on the main board and on PI-bus input on the Decagon interface.
The wiring to RS485 follows the labelled colors on the pcb, while on the Decagon interface is the following: WH=brown, RD=green, BR= white.
DIP switch for PI-bus is 111.
3.2 INSTALLING SOIL MOISTURE SENSORS
iMETOS® weather and soil moisture stations support awide range of soil moisture sensors. This manual covers the most common of them.
In particular, iMETOS can be equipped with 4 different types of soil moisture sensors. Tensiometers are giving accurate readings in between 10 and 700 mbar of water tension. Tensiometers for this are widely used in horticulture because their measurement range fits very well with the most vegetables and soft fruits. Watermark sensors are returning water tension too. The range of this sensor is between 10 and 2000 mbar. This sensor fits for all horticulture crops, soft fruits, for all tree fruits and for all agricultural crops. It is easy to use and allows applying a controlled deficit regime for crops that can accept this.
Then we have capacitive probes which give the amount of water what has been used by the crop but it tells nothing about the availability of the water. Sentek Drill&Drop or Aquacheck probes are giving a perfect way to monitor the water use in different soil depths.
On iMETOS we can combine tensiometers with watermark sensors, capacitive and profile probes with the use of nodes and interfaces. The combination of Tensiometer and Watermark will give accurate readings by the tensiometer in the area when water is easily available for the plant and readings for controlled water deficit when this is applied. Please note the tensiometers should be removed in the time of water deficit or refilled after this time. The combination of tensiometer with capacitive probes will give information about the availability of soil water and about the water needed to refill the soil.
Please note that before connecting a NEW INTERFACE to the motherboard, delete sensor configuration and unplug battery and solar panel. Once connected, plug them in again.
It is recommended to do always a SENSOR TEST before leaving the site to ensure the station reads properly the new sensor.
From tensiometric soil moisture measuring they support:
– Tensiometers with Pessl Instruments pressure transducer head
– Watermark sensors
– METER Group MPS
From volumetric water content measuring they support:
– Sentek Drill&Drop
– METER Group 10HS
– METER Group EC5
– METER Group 5TE
– METER Group 5TM
– METER Group GS1
– METER Group GS3
– METER Group MPS1
The soil moisture sensors are supported by nodes for PI sensor bus. These nodes connect PI sensor bus to RS485 input on the PCB. iMETOS® 3.3 has one native RS485 input and can be expanded by 2 more on the extension connector. The PI sensor nodes come with a shielded 4-wire cable.
The cable has the following color code:
– Bare, Gray or Blue: Shield
– White: Ground
– Yellow: Data A
– Green: Data B
– Brown: VVC
The RS485 connectors on the PCB are labelled with SH(D) (Shield), WH(T) (Ground), YE(L) (Data A), GR(N) (Data cool and BR(N) (VVC). With a little interface PCB in the cable all the sensor bus nodes can be connected to the inputs 1 to 4 too.
To install fork-like sensors, excavate a trench to the depth required and insert the entire sensing portion of the sensor horizontally into the undisturbed soil face. Let the cable go down first and then take it up to the top. This prevents water from following the cable in the ground and wetting the sensor.

3.3 INSTALLING TENSIOMETERS
Tensiometers are returning water tension in cBar, kPas or mBar. The pressure would be a negative value because it indicates tension, but for simplification FieldClimate handles them as positive.
Tensiometer sensors are supported by the PI-bus. Every pressure transducer comes with a shielded 3 wire cable. The wires and the labels on the PCB are: Bare: SH, WH: white, GR: green, BR: brown. The connection is the following: Brown wire on WH, Green on RD, White on BR.
The tensiometers are having an outside diameter of 20mm. Use a soil auger to make a narrow hole and moisture the hole with water before you enter the tensiometer. The tensiometer has to have close contact on the side and at the base of the hole. Water tension measurement is very sensible to air gaps below the sensor.
Tensiometers are returning water tension in cBar, kPas or mBar. The pressure would be a negative value because it indicates tension, but for simplification FieldClimate handles them as positive.
Tensiometer sensors are supported by the PI sensor nodes. There is a sensor node supporting 4 tensiometer pressure transducers and having a reference pressure sensor on the PCB. The inputs are
enumerated from 1 to 4. On FieldClimate the data is displayed following this enumeration with the atmospheric pressure from the reference sensor as the leading value.
Every pressure transducer comes with a shielded 3 wire cable. The wires and th labels on the PCB are:
Bare: SH, WH: white, GR: green, BR: brown.
3.3.1 FILLING AND RE-FILLING THE TENSIOMETER
To re-fill the tensiometer please emerge the sensor totally overnight in water filled bucket. The next morning please connect a syringe to the tensiometer and pull it up. Use the syringe to degas the tensiometer. When the syringe fills with water remove it and plug the tube again.
3.4 INSTALLING WATERMARK SENSORS
Watermark sensors measure water tension. These sensors measuring range is between 100 and 2000 mBar (10 to 250 kPa or cB). This sensor fits many horticultural crops, berries, stone fruits, vines and potatoes. Watermark sensors are easy to use and frost-proof.
Basically it is a gypsum block. The gypsum is embedded in a matrix material and gives a defined response in resistance on different water tension levels. We can measure the water tension on 3 to 5 cBar accuracy.
It showed that it works faster if it is wetted for 24 hours before installation. If you install it in the ground you will need a hole with the correct depth. Lay it horizontally on the ground and let the cable go horizontally first to before you take it up to the top. This avoids water following the cable in the ground, thus reaching and wetting the sensor.
The watermark PCB can be equipped with an own soil temperature sensor. This sensor is used to do the temperature compensation of the water tension measurement. It allows you to to measure the soil temperature in specific blocks where you measure water tension. The irrigation might influence the soil temperature what can be important in asparagus for example.
3.5 INSTALLING DRILL&DROP PROBES
Up to two Sentek “Drill And Drop” volumetric water content in the profile probes can be connected via extension board directly connected to the main PCB.
Further number od this probes can be connected via PI bus input or via radio node links. Refer to D&D manual for more details.



3.6 WIRING A JOHN DEERE PROBE
- Install the interface board SDI-12 in the IMT/ECO D3
- Run the wires out one of the wire plugs
- Make sure the wires are firmly installed


3.7 WIRING AN AQUACHECK PROBE
- Instruction for wiring the Aquacheck soil moisture probe to METOS SDI 12 interface
- Install the interface board
- Run the wires out one of the wire plugs
- Make sure the wires are firmly installed
- Connect the Blue wire from the Aquacheck probe to the terminal marked DAT.
- Connect the Yellow/Green stripe wire from the Aquacheck probe to the terminal marked GND.
- Connect the Brown wire from the Aquacheck probe to the terminal marked PWR.


3.8 iMETOS® 3.3 RADIO ACCESS POINT
iMETOS® iMETOS 3.3 includes the chipset for radio linking remote radio-nodes of sensors and actuators. The radio-node net has a star topology (iMETOS® controls the net, RadioNodes are peer to it). 16 remote radio-nodes can be connected to the iMETOS® 3.3. Radio-nodes should be places in a line of sight at a maximum distance of 500m from the iMETOS®.

3.9 SETUP YOUR iMETOS® WiFi NETWORK
Before you can start using your iMETOS® WiFi you need to correctly setup the connectivity settings, that your device will be able to connect to your local WiFi network. To do this, please follow the following steps:
1. Start the station with jumper on J1 position and connect it with your. When this is done please enter the terminal window program and plug in the battery.
2. In terminal window program, you will see basic menú, please select option nr. 3 – MODEM.
========================================================================================
USER: \ MAIN MENU
========================================================================================
(1) SYSTEM
(2) SENSORS
(3) MODEM
3. Select (1) MODEM SETTINGS and please wait as station checks the modem.
29/03/2020 00:20:42 >> Chosen function: MODEM
========================================================================================
USER: \ MAIN MENU \ MODEM
========================================================================================
(1) MODEM SETTINGS
(3) SEND STATION DATA
(4) GET GPS
(5) GET CONF SMS
(6) Download FW
(7) Download APN list
(ESC) BACK
29/03/2020 00:20:45 >> Chosen function: MODEM SETTINGS
>
> Starting modem process
>
>
>> Turning modem on
>
Try saved baudrate: 38400
ATE1
[ERROR: INVALID INPUT]
ATE1
[OK]
Baudrate detected
>
>> Reading modem’s info
>
AT+NMAC=?
00:08:dc:1c:9d:62
[OK]
ATI1
GS1011
[OK]
ATI0
WIZnet
[OK]
ATI2
WizFi210
1.1.1.0(W)
[OK]
Manufacturer of modem: WIZnet
Type of modem: GS1011
Modem firmware version: WizFi210__1.1.1.0(W)
MAC: 00:08:dc:1c:9d:62
Modem Baudrate: 38400
========================================================================================
USER: \ MAIN MENU \ MODEM \ MODEM SETTINGS
========================================================================================
(1) SEE AVAILABLE NETWORKS
(2) CONFIGURE NETWORK
(3) DISPLAY NETWORK CONFIGURATION
(4) ERASE NETWORK CONFIGURATION
(5) TEST CONNECTION
(ESC) BACK
4. In the menu, select option nr. 1 – See available networks. Please find suitable network on the list. Please take a look at RSSI number – the bigger the number is the better is signal on the location (note that it is a negative value – so in this case the best is -57).
29/03/2020 00:21:10 >> Chosen function: SEE AVAILABLE NETWORKS
========================================================================================
USER: \ MAIN MENU \ MODEM \ MODEM SETTINGS \ SEE AVAILABLE NETWORKS
========================================================================================
Scanning, please wait…
BSSID SSID Channel Type RSSI Security
46:d9:e7:cd:55:f3, Pessl WiFi, 01, INFRA , -57 , WPA2-PERSONAL
00:1f:33:fb:b5:26, Pessl_Dev, 06, INFRA , -86 , NONE
46:d9:e7:cd:53:30, Pessl WiFi, 06, INFRA , -80 , WPA2-PERSONAL
46:d9:e7:cd:55:11, Pessl WiFi , 11, INFRA , -76 , WPA2-PERSONAL
No.Of AP Found:4
5. Now navigate to menu nr. 2 – Configure network. Now please follow the on-screen instructions. Input the right network name, type of security and password. Usually Automatic IP and Automatic DNS is set to YES.
========================================================================================
USER: \ MAIN MENU \ MODEM \ MODEM SETTINGS
========================================================================================
(1) SEE AVAILABLE NETWORKS
(2) CONFIGURE NETWORK
(3) DISPLAY NETWORK CONFIGURATION
(4) ERASE NETWORK CONFIGURATION
(5) TEST CONNECTION
(ESC) BACK
29/03/2020 00:46:28 >> Chosen function: CONFIGURE NETWORK
========================================================================================
USER: \ MAIN MENU \ MODEM \ MODEM SETTINGS \ CONFIGURE NETWORK
========================================================================================
Please select the proper security option:
(1) WEP (KEY 1)
(2) WPA
(3) WPA2-PSK (Using Password)
(4) WPA2-PSK (Using PSK)
(5) NONE
(ESC) BACK
and now set your Network Parameters.
6. After successfully inputting new settings for WiFi network, you can check on menu nr. 3 – Display network configuration, if all is as you set it up.
7. As the last step you should run a test, so in the menu please choose nr. 5 – Test connection. You will see the full list of all available networks, and then the station will try to connect to the network of which you provided the settings. If at the end of this process there is a massage: INFO: Network properly configured, then everything is ok and you can start using your iMETOS® station. If you get a massage ERROR: Network NOT properly configured, then the provided configuration is not working. In this case, please make sure that you have entered the right parameters.
29/03/2020 00:47:44 >> Chosen function: TEST CONNECTION
========================================================================================
USER: \ MAIN MENU \ MODEM \ MODEM SETTINGS \ TEST CONNECTION
========================================================================================
ATC0
[OK]
AT+WM=0
[OK]
AT+WS=iMetosWiFi
BSSID SSID Channel Type RSSI Security
46:d9:e7:cd:55:f3, Pessl WiFi,01, INFRA,-57,WPA2-PERSONAL
No.Of AP Found:1
[OK]
> Chosen network: BSSID d4:28:d5:5e:90:f3 with RSSI: -91
AT+WPAPSK=iMetosWiFi,wifi1234
Computing PSK from SSID and PassPhrase…
[OK]
AT+NDHCP=1
[OK]
AT+WA=iMetosWiFi,d4:28:d5:5e:90:f3
[ERROR]
AT+WD
[OK]
4. INSTALLING NEW SENSORS ON THE iMETOS®
4.1 INSTALLING AIR, SOIL, WET OR DRY BULB TEMPERATURE 4.2 INSTALLING A GLOBAL RADIATION SENSOR 4.3 INSTALLING A WIND SENSOR4.4 INSTALLING BAROMETRIC PRESSURE SENSOR
Each iMETOS® can be expanded to the maximum possible number of sensors. It is equipped with specific inputs for rain, wind speed, leaf wetness, temperature and relative humidity and for watermarks, gypsum blocks or echo probes. Beside of this it has 3 digital inputs more which can be used for a range of different sensors and sensor chains.
iMETOS® is able to distinguish in between the different sensors added to the 3 digital inputs. If we connect a global radiation, a photosynthetic active radiant sensor, a wind direction, a barometric pressure sensor or a temperature sensor – they can be detected by their frequency. When we connect digital sensors with a numerical output like chains of tensiometers, they can be detected by their identifier.
4.1 INSTALLING AIR, SOIL, WET OR DRY BULB TEMPERATURE
iMETOS® is able to distinguish between different sensors. All the temperature sensors, that iMETOS® is normally equipped with, are basing on the same IC and they are not distinguishable. Therefore iMETOS® make assumptions on the most common temperature sensors for horticulture and agriculture.
If you connected the temperatures to these connectors and they are not one of the sensors which are preset, please rename them on the website FieldClimate -> Settings -Y Sensors and Nodes. If you are going to rename the sensors in the internet, every combination of temperature sensors is possible.

4.2 INSTALLING A GLOBAL RADIATION SENSOR
The global radiation sensor identifies itself by its frequency. It can be connected to any input from 1 to 4. The design of the iMETOS® makes it more comfortable to enter the box from the right side and to use the input 1.
To fix the global radiation sensor on the iMETOS® holder unscrew the three screws right of the box and fix the global radiation holder with this screws.
4.3 INSTALLING A WIND SENSOR
4.3. 1 PI ULTRASONIC WIND MONITOR 4.3. 2 WIND DIRECTION SENSOR
4.3. 1 PI ULTRASONIC WIND MONITOR
PI Ultrasonic Wind monitor measures accurately wind speed and direction.
The sensor is to be mounted at the top of the second pole of the iMetos and with the standard clamps. It is supplied with a 2-meter PIbus cable (optional is to include the longer cable when placing your order). It gives the following values:
- Average wind speed in the logging interval
- The maximum gust of wind in a running 3-seconds average
- Wind run direction
You will take the same installation care and placement considerations as for the traditional anemometer with the plus of a careful orientation to the north pole guided by the “north mark” indication in the sensor.
It measures 3 times per second and applies a median-filtering to save this value in its embedded logger. On request of the iMetos, it performs the averaging and the pertinent vectorial calculations to supply the 3 mentioned values.
It is powered with a small super-cap charged in milliseconds by the iMetos during every reading. The power uptake is negligible. It requires no maintenance and fulfills and improves all the measurement demands.
Scheme connection on iMETOS 3.3
Pancon connector: Input 1, input 2, input 3, input 4.

In case you need to connect it to PI-bus RS485, you have to cut and strip the cable before the black heat shrink tubing, which contains the RS485 converter.
Connection to RS485 on the main board with the following wiring color: BR=brown, GR=green, YE=yellow, WH=white.
To install the PI Ultrasonic wind sensor on the 2nd pole (42.4 mm) for iMetos 3.3 follow the next steps:
- Bend the cable about 10 cm from the sensor.

- Place the Ultrasonic sensor on the top of the second mounting pole. The PI Ultrasonic Wind is fitted with two clamps and one metal holder, allowing it to be fixed on the pole or post.
- The PI Ultrasonic Wind must be oriented with the „North mark“ sticker below the sensor to true north. Check it with the usage of Compass mobile app.
- Use a wrench to tighten the bolts and let the cable passing through the screw and the pole (see figure below).

- Fix the cable vertically along the pole with zip ties avoiding to wrap it around the pole.
4.3. 2 WIND DIRECTION SENSOR
The wind direction sensor identifies itself by is frequency. It can be connected to any input from 1 to 4. The design of the iMETOS® makes it more comfortable to enter the box from the downside and to connect the sensor to input 2 or to input 4.
The wind direction sensor is different from the other sensors in its mode of measurement. It uses its own intelligence to measure every second. It transmits the average of its readings on every request from the main unit. To be able to measure during the periods, the main unit sleeps, this sensor has its own small battery. This battery should last for 10 years. If this sensor fails to work after several years, please note that it might be the battery which is empty.
On northern hemisphere the wind direction sensor should point towards north and the wind speed sensor should point towards south again.

4.4 INSTALLING BAROMETRIC PRESSURE SENSOR
The barometric pressure sensor can be connected to one of the digital Inputs 1-4 on the PCB inside the box. To install it, attach it with the tape (it is included) on the metal holder for the iMetos next to the box containing the pcb. The following pictures show the steps for the installation of the sensor.


5. MAINTAINING THE iMETOS®
5.1 MAINTAINING THE RAIN GAUGE 5.2 MAINTAINING THE LEAF WETNESS SENSOR 5.3 MAINTAINING THE TEMPERATURE AND RELATIVE HUMIDITY SENSOR 5.4 MAINTAINING THE SOIL TEMPERATURE SENSOR 5.5 MAINTAINING THE GLOBAL RADIATION SENSOR 5.6 MAINTAINING THE WIND SPEED SENSOR
When the solar panel of the iMETOS® is exposed to the sun and it gets enough sunlight it should recharge the lead acid battery of the system constantly. With this the system should have enough power for years. The lifespan of the lead acid battery can be expected to be 5 to 6 years with sufficient reloading from the solar panel. Insufficient reloading will shorten the lifetime of the battery. In case of insufficient reloading of the battery the system will reduce the frequency of data upload to the amount which is needed to make sure that no data will be missed on the server. With this, power use will be reduced, and battery power will last longer.
The maintenance needed for the sensors depends on the different sensors. Temperature sensors will not need any maintenance for their lifetime whereas the rain gauge will only work properly when it is cleaned periodically.
5.1 MAINTAINING THE RAIN GAUGE
New rain-gauge sensor funnel includes a plastic filter intended to reduce maintenance needs. It is possible that birds or insects construct nests inside the mechanism; so, if it rains and the data does not look consistent, check your Rain Gauge internally. To open it, just press the lateral walls and pull the metallic lid (see pictures below).


5.2 MAINTAINING THE LEAF WETNESS SENSOR
The filter paper which is used as a sensing element for the leaf wetness sensor is destroyed after to many wettings and dryings. It has to be changed in dependence of rain frequency.
5.3 MAINTAINING THE TEMPERATURE AND RELATIVE HUMIDITY SENSOR
The temperature sensors of iMETOS® are free of any maintenance as long as they are used in a proper way and the cables are not damaged. The relative humidity sensor is affected by pollution in dependence of the severity of the pollution. In clean air this sensor will give accurate readings for 3 and more years but in polluted air it may give inaccurate high readings after some years of use. In this case the sensor has to be changed for a new calibrated one.
5.4 MAINTAINING THE SOIL TEMPERATURE SENSOR
The soil temperature sensor does not need any maintenance. If it fails it is nearly always due to damaged wires.
5.5 MAINTAINING THE GLOBAL RADIATION SENSOR
The global radiation sensor should be cleaned every month. After 2 or 3 years the sensor has to be calibrated in our factory.
5.6 MAINTAINING THE WIND SPEED SENSOR
The wind sensor should be examined whether the windwheel is low-friction. If that is not the case sent the sensor for sevice to us. If the windwheel is damaged please order a new one.
6. FIRMWARE
6.1 UPDATE OVER THE AIR 6.2 STATION WORKING MODES 6.3 EVENTS
6.1 UPDATE OVER THE AIR
The firmware is the application that runs on the processor of the iMETOS®. Every time the iMETOS® 3.3 connects to FieldClimate, it checks for the latest firmware version. If it finds a newer version the iMETOS® 3.3 automatically downloads it and updates its self.
In addition firmware can also be uploaded manually via the USB connection. A detailed description is given on pages about USB communication.
6.2 STATION WORKING MODES
The iMETOS® 3.3 can operate in three different working modes:
1. Logging and transmitting (normal mode)
2. Logging, not transmitting (gathering mode)
3. No logging, no transmitting (sleeping mode)
The standard working mode is the normal mode in which the station measures and communicates regularly.
When battery voltage drops below a threshold (6.3V), the fixed transfer interval is ignored.
If the battery voltage drops below a threshold (6.1V), the station changes to gathering mode. The station measures regularly, but modem communication is deactivated.
If the battery voltage falls below yet another threshold (5.4V), the station changes to sleeping mode. Both measurements and communication are deactivated.
A station in gathering mode automatically changes back to normal mode if the voltage is high enough again.
If the station is in sleeping mode, also the measurement of the battery voltage is deactivated. Therefore the station has to be activated manually by pressing the connect button (see Picture 29) on the board in order to change back to normal mode. First the SMS are checked, then the communication with the server is started.

6.3 EVENTS
Along with the weather data, the iMETOS® 3.3 also sends relevant info concerning previous communication, SMS readings, internal parameters and many other things.
User can see this events at FieldClimate (Fieldclimate.com / Settings / Station info / Events details) or locally via USB connection.
Please refer to appendix 2 for the complete list, here as an example:

7. WEB SERVER COMMUNICATION
7.1 STATION TO WEB-SERVER COMUNICATION 7.2 WEB-SERVER RESPONSE 7.3 USER DEFINED SETTINGS
7.1 STATION TO WEB-SERVER COMUNICATION
Data from station to the server is sent in XML format.
The XML file contains:
• station info (e.g. firmware version, serial number,…)
• the station settings (e.g. rain monitoring settings, measurement intervals,..),
• modem info and GSM settings (SIM ID, APN settings,…),
• the events since the last communication,
• the measured data (“weather bin”, encoded in Base64 format).
7.2 WEB-SERVER RESPONSE
After receiving the station data, the web-server parses the data, stores it in the database and sends a response (also in XML format). In this response the newest versions of firmware, APN table and sensor table are transferred, the settings are updated and the success of the database upload is confirmed. In detail this includes:
1. IM-RTC-TIME: date and time for synchronization
2. IM-STATION-NAME: user-defined station name
3. IM-UOTA: available firmware version to download
4. IM-FWCRC: firmware correction value
5. IM-FWFILE: name of firmware binary file
6. IM-APNTABLE: available APN table version
7. IM-SENSTABLE: available sensor table version
8. IM-MODE: station mode
9. IM-WL-MON: user-defined water level monitoring settings
10. IM-RAIN-ENABLED: rain monitoring enabled flag
11. IM-MEAS-INT: user-defined measurement and logging intervals
12. IM-FIXED-TRANSF-INT: user-defined fixed transfer interval
13. IM-RAIN-INT: user-defined rain monitoring interval
14. IM-WL-INT: user-defined water level monitoring interval
15. IM-EMG-SMS: user-defined emergency phone number
16. IM-DATA-SCHED: user-defined data transfer schedule
17. IM-SMS-WARN-GSM: user-defined SMS warning phone number
18. IM-SMS-WARN-VAL: user-defined SMS warning sensor settings
19. IM-DB-DATE: begin and end date of data stored at server side
20. IM-DB-UPLOAD: Database upload success flag (if 1 successful, if 0 not successful)
* this settings can change in future
7.3 USER DEFINED SETTINGS
7.3. 1 RAIN MONITORING 7.3. 2 WATER LEVEL MONITORING 7.3. 3 MEASUREMENT AND LOGGING INTERVAL 7.3. 4 FIXED TRANSFER INTERVAL AND TRANSFER SCHEDULE 7.3. 5 EMERGENCY SMS 7.3. 6 SMS WARNING SYSTEM
In this section the user-defined settings are given. They can be configured for each station in FieldClimate.
7.3. 1 RAIN MONITORING
The user can define a rain monitoring interval. If rain monitoring is enabled, the iMETOS® 3.3 measures, logs and transfers data in the specified interval while it is raining. By default rain monitoring is off.
This option is available only at the stations, which have rain sensor.
If the measurement and logging interval are smaller than the specified rain monitoring interval, they are kept.
Note:
iMETOS® 3.3 supports rain-gauges with different resolutions, (i.e.: 0.1mm, 0.2mm, 0.5mm) user sets the corresponding resolution in the FieldClimate portal settings. User can also enable pulse filtering when it is necessary to ignore fast pulsing due to vibrations or wind.
For SMS rain alarms a second rain counter accumulates the rain when it is in the same rain event independently of the timestamps.
7.3. 2 WATER LEVEL MONITORING
The user can define a water level monitoring interval and a water level threshold. When the threshold is exceeded, the iMETOS® 3.3 measures, logs and transfers data in the specified interval.
If the measurement and logging interval are smaller than the specified water level monitoring interval, they are kept.
7.3. 3 MEASUREMENT AND LOGGING INTERVAL
The measurement interval defines, when a sensor does measurements. The default measurement interval is 5 minutes.
The logging interval defines when the measured data is logged (meaning: measured data is combined into one block and the minimum, maximum, average,… values of each sensor within this logging interval are calculated). The default logging interval is 60 minutes.
Before each transfer, an additional logging event is triggered.
Each sensor chain (i.e. each input on the iMETOS® 3.3 board) may have distinct measurement and logging intervals.
7.3. 4 FIXED TRANSFER INTERVAL AND TRANSFER SCHEDULE
The times when an iMETOS® 3.3 attempts to communicate with the webserver and transfer the logged data can be defined in three ways:
1. A user-defined data transfer schedule: the user can define transfer times at full hours on a weekly basis,
2. A user-defined fixed transfer interval,
3. A user-defined forced transfer interval, which is the same as the fixed transfer interval but limited to a certain timespan.
Default is a fixed transfer interval of 60 minutes. Web-server communication and data transfer are also started due to the following reasons:
1. If rain monitoring is enabled, the iMETOS® 3.3 sends data to the server every specified interval during precipitation events,
2. If water level monitoring is enabled, the iMETOS® 3.3 sends data to the server every specified interval when the water level is higher than the threshold,
3. If the Connect button on the board is pressed.
7.3. 5 EMERGENCY SMS
The user can define one phone number, to which the iMETOS® 3.3 sends out an emergency SMS if the battery voltage falls below a certain threshold.
7.3. 6 SMS WARNING SYSTEM
The user can define an upper and lower threshold SMS alarm values for each sensor. If a measured value exceeds (or goes below) the establish threshold, the iMETOS® 3.3 will send the SMS with a warning text (name of sensor, actual value, threshold value) to up to 10 different mobile phone numbers.
Alarm thresholds are checked every 5 minutes within each measurement. When the iMETOS® receives the sent acknowledgment from the mobile network the iMETO®S 3.3 will disable SMS sending for 4 hours.
But if a measured value drops below the threshold in 3 consequent measurement cycles, the iMETOS® will again enable the sending of SMS and send the corresponding alarm.
Example:
Temperature sensor threshold for SMS warning is set to 30°C
Measurement 1: 9:00 value = 28°C
Measurement 2: 9:05 value = 30°C -> a warning SMS is send
Measurement 3: 9:10 value = 31°C -> within a 4hour block interval -> warning SMS is not send
Measurement 4: 9:15 value = 28°C
Measurement 5: 9:20 value = 29.5°C
Measurement 6: 9.25 value = 29°C -> the blocking interval is cleared
Measurement 7: 9:30 value = 31°C -> warning SMS is send and block interval is started again
After each SMS is sent the iMETOS® 3.3 will maintain the modem “on” and establish a communication with FieldClimate to upload the complete set of stored data since the previous communication and to receive new settings (if there are any) from the user.
8. SETTING COMMUNICATION PARAMETERS AND RESETTING iMETOS® VIA SMS
A user-defined APN can be set via SMS when the required information is not yet in the APN table of the iMETOS®. Also a station reset via SMS is possible.
Notes:
• You need the phone number associated to the SIM card in the iMETOS® 3.3
• Alternatively, to send the SMS to the SIM card at the iMETOS® 3.3 you can use a SIM card which already has the configuration SMS locally stored.
• After sending the SMS described below in this section press the connect button in the iMETOS® so it checks for received SMS!
The following commands can be send via SMS to the iMETOS® 3.3:
1. Set user-defined APN:
Code:! serial_number 0 apn,user_name,password !
serial_number is the serial number of the station,
apn is the APN server address,
user_name is the user name of APN server,
password is the password of APN server.
Example: ! 00200003 0 a1.net,gprs,a1!
Note: there is no space between the APN parameters but there is one space between exclamation mark, serial number, SMS parameter, APN string and exclamation mark.

An APN send via SMS is stored in the internal APN XML file, and if correct, is set as default.
If the APN does not work, the system will try to find another APN access from the APN table and the user-defined APN will not be used anymore with the SIM. In order to force the system to retry using a user-defined APN, change the SIM card (as this forces the system to again find a correct APN in the whole APN XML file) or send a new SMS.
To reset the user-defined APN send an SMS without parameters: ! serial_number 0 ,, !
2.Set user-defined APN full definition:
Code: ! serial_number 3 country_name,MCC,MNC,apn,user_name,password !
serial_number is the serial number of the station
country_name is the country of the provider
MCC is the MCC of the provider
MNC is the MNC of the provider
apn is the APN server
user_name is the user name of APN server
password is the password of APN server
Examples: 00202233 3 austria,232,1,a1.net,gprs,a1 !
argonaut 3 spain, 214,07,movistar.es,movistar,movistar !
Note: there is no space between the APN parameters but that there is one space between exclamation mark, serial number, SMS parameter, APN string and exclamation mark.

3.Reset station:
Code: ! serial_number 2 UID_station !
serial_number is the serial number of the station
UID_station is the Unique identification number of the station
You can get the UID_number by connecting the iMETOS® to the PC via USB cable and asking for a quick report (see pages abour USB communication).
This command does a factory reset (exception: the APN table is kept).
9. USB COMMUNICATION (PI FIRMWARE UPLOADER)
9.1 USB COMMUNICATIONS MODES 9.2 USING TERMINAL MODE 9.3 TESTING SENSORS 9.4 UPLOADING FIRMWARE 9.5 CHANGING THE SERIAL NUMBER
Direct communication with the iMETOS® 3.3 is possible via a standard mini USB cable that is connected to the USB connector. The standard interface for communication is the “iMETOS® Firmware Uploader”.
Requirements:
•.NET 3.5 or higher
•EFM32-cdc USB drivers have to be installed
•PI Firmware Uploader has to be installed
The iMETOS® Firmware Uploader and the appropriate drivers can be downloaded at: /redmine/projects/pi-firmware-uploader/files
Alternatively contact our support team at support@metos.at
9.1 USB COMMUNICATIONS MODES
Depending on the jumper configuration on the iMETOS® 3.3 board one of the following USB communication modes is active:
•Terminal mode (Jumper 1 set – G in the picture below):
In terminal mode active communication with the iMETOS® 3.3 is possible.
•Modem direct communication mode (Jumper 2 set – F in the Picture):
Direct active communication with the modem is possible.
•Boot mode (Boot jumper set – J in the Picture):
<>In boot mode new firmware versions can be uploaded manually.
•Spy mode (no jumper set but USB cable is connected):
In spy mode the iMETOS® 3.3 carries out the normal processes like measuring and communicating with the web server. Real-time information on the processes is printed
to the screen and can be used for monitoring, testing and error tracing.


1. Place the appropriate jumper and plug in the USB.
2. If the battery is unplugged: Plug in the battery.
3. If station is in boot mode: Press the Reset button on the iMETOS® 3.3 board to exit boot mode. Check if the jumper is not set on boot pins.
4. In PI Firmware Uploader: Click “Open Port” button.
5. Communicate with station.
6. Optional: Switch between normal, terminal and modem direct communication mode by changing the jumper position.
7. Click button “Close Port” when you want to end the communication.
8. Unplug USB.
The most common application of the boot mode is to manually upload a new firmware version.
Possible reasons for the error message “COM port unavailable” or similar:
• The board is not connected to the power supply.
• The board was reset while the port was open.
Solution: Close Port and reset the board again by pressing the Reset button. If needed, try to unplug and plug back again the USB cable.
9.2 USING THE TERMINAL MODE
With the battery connected to the iMETOS® 3.3 proceed as follows:
1. Plug USB cable to the iMETOS® 3.3
2. Plug the USB cable to the PC
3. Place Jumper 1
4. Run the application PI_uploader
5. Click on “Open Port” button You will get the main menu. If not press space bar in your keyboard

=======
MAIN MENU
=======
(1) SYSTEM
(2) SENSORS
(3) MODEM SETTINGS
6. Press 1 to go to the system menu and get:
=======
MAIN MENU > SYSTEM
=======
(1) QUICK VIEW
(2) FULL REPORT
(3) FACTORY RESET
The following options are available:
(1) in the menu SYSTEM:
(1) QUICK VIEW: gives some basic station, modem and server information.
(2) FULL REPORT: gives a detailed report on the station including the station settings and events.
(3) FACTORY RESET: discards the station settings, measured data and APN table.
7. press Esc to return to main menu
8. press 2 to get sensors menu
=======
MAIN MENU > SENSORS
=======
(T) DO SENSOR TEST
(W) WIRELESS TEST
(P) PRINT LAST MEASURED DATA
(S) PRINT LIST OF SUPPORTED SENSORS
(I) STORE MEASURED DATA
(R) PRINT ENCODED RECORDS
(D) DISCARD SENSOR DATABASE
(ESC) BACK
(2) in the menu SENSORS:
(T) TEST OF SENSORS: opens the sensor test menu (see below).
(M) DO MEASUREMENT ONLY: prompts the iMETOS® 3.3 to start a measurement. A submenu is opened in which you can select a chain by pressing (1) – (7) or all sensors by pressing (A).
(P) PRINT LAST MEASURED DATA
(S) PRINT LIST OF SUPPORTED SENSORS
(I) STORE DATA IN MEM: prompts a logging event.
(R) PRINT ENCODED RECORDS: shows all stored data encoded as weather bin (data transfer format)
(D) SET DATABASE TO DEFAULT: discards the station settings and measured data (the APN table is kept).
9. press Esc to return to main menu
10. press 3 to get the modem settings menu
=======
MAIN MENU > MODEM SETTINGS
=======
(1) GET MODEM INFO
(2) UNLOCK SIM CARD (PIN CODE)
(3) SIGNAL QUALITY GRAPH
(4) APN OF CUSTOMER
(5) SERVER REDIRECTION
(3) in the menu MODEM SETTINGS:
(1) GET MODEM INFO: shows IMEI, SIM card number, type of modem…
=======
MAIN MENU > MODEM SETTINGS > GET MODEM INFO
=======
Modem Brand: Sierra Wireless
Modem Type: SL6087 Product
Modem FwVersion: R7.46
Modem IMEI: 354293068370835
Modem bearer type: GPRS
Modem Baudrate: 115200
SIM status: Inserted ( SIM ID: 8943015614107200408 )
SIM locking: UNBLOCKED
(2) UNLOCK SIM CARD: check whether the PIN number request is enable and also provides means to disable it.
(3) SIGNAL QUALITY GRAPH: gives a representation of the carrier signal quality
=======
MAIN MENU > MODEM SETTINGS > SIGNAL QUALITY GRAPH
=======
—
(Esc) <- exit from this menu item – Signal: >| | | | | | | | | …< ( RSSI: 29 / BER: 0 )
(4) APN OF CUSTOMER : it let you directly configure the Access Point Name for the specific provider
=======
MAIN MENU > MODEM SETTINGS > APN OF CUSTOMER
=======
CURRENT APN ACCESS DEFINED BY USER:
-There is not defined APN by user
CHOOSE THE ACTION:
– –
(I) INSERT NEW APN ACCESS
(D) DELETE DEFINED USER APN
(5) SERVER REDIRECTION: It allows you to change the ULR address of the server
=======
MAIN MENU > MODEM SETTINGS > SERVER REDIRECTION
=======
CURRENT SERVER SETTINGS:
Domain name: metos.at
URL address: /pikernel_dev/metos_upload_xml33.php
Port: 80
CHOOSE THE ACTION:
– –
(R) REDIRECT SERVER
(D) RESET SERVER REDIRECTION
Clue: If there is no menu visible, press Space on your keyboard and the actual menu will be shown. With ESC parent menu will be shown.
Note: Options in the Terminal Menu can be changed with new Firmware.
9.3 TESTING SENSORS
To test the sensors start terminal mode and navigate to the Test of sensors menu as described above. There you can select the chain to test by pressing (1) – (8) or you can test all sensors at once by pressing “A” on your keyboard. While the test measurement is in progress, the yellow LED is turned on. This may take several minutes. The output then printed on the screen has the format:

You can repeat the measurement by pressing “r” or quit the test by pressing “q”.
9.4 UPLOADING FIRMWARE
Eventually it can be necessary to manually upload firmware. To do this follow the given steps:
1. Plug the USB and set the Boot jumper. If not done yet, insert the SIM card and connect the antenna.
2. If the battery is unplugged: Plug battery.
3. Check whether the station is in boot mode and the Pessl bootloader is installed: Yellow LED in the lower row is on.
4. If the station is not in boot mode: Press Reset button on the board.
5. In Firmware Up-loader:
(1). Click “Select FW” and select the correct firmware binary file.
(2). Check that Device “iMETOS® 3.3” is selected.
(3). Select the correct COM Port.
(4). Click “Upload firmware”.
(5). After successful upload, the board automatically restarts, starts communication with the web-server and enters the boot menu. The boot menu is a simple version of the terminal menu, offering only limited options.
(6). To leave the boot menu/boot mode click “Close Port”, remove the Boot jumper, remove the USB cable and press reset on the board.
9.5 CHANGING THE SERIAL NUMBER
In order to change the serial number of a station, an *.ini file containing the new serial number is needed. *.ini files are issued by mail upon request (support@metos.at).

If you received your *.ini file, copy it to the path where the firmware uploader application is installed and start the firmware uploader. The new serial number is recognized automatically. Follow the instructions to upload firmware given above. After the upload, the *.ini file is deleted automatically!
Important note:
Complete monitoring of the communication process should be done after every installation by connecting the PC to the iMETOS®. Installer never should leave the site without performing sensor test and communication process log.
10. APENDIX I.
10.1 SENSOR CODES
Up to the date of this edition, a total of 182 codes for different types of connectible sensors has been defined for the iMETOS® 3.3. Continuous development further increases it.
11. APENDIX II.
Up to this date there are a total of 182 codes of this edition that are defined for the iMETOS® 3.3. These codes include 182 for different types of connectible sensors. Continuous developing from our side increases this list constantly.
11.1 LIST OF EVENTS